Liquid Nitrogen LN2 Safety

LN2 has a boiling point of -195C. If it touches your skin, it will severely burn you. LN2 expands almost 700 times in air. It can deprive a closed space from oxygen, leading to death of anybody in the room.

Video LN2

As long as you enter the lab, you have the oxygen monitor turned on and on you. It is especially important to do so when you are using liquid nitrogen to test, because during a test, LN2 is poured into the chamber and evacuated directly in the room.

Cryogenic safety training

You have to complete the cryogenic safety training to be allowed to work with liquid nitrogen. You can also contact Tom Benassi in the ME office if you need hands-on training and nobody in the lab is able to provide it.

You should enter the number for JHU Security in your phone so that if there is an emergency and you need to leave the room (generally first thing to do), you will be able to call JHU Security from the hallway.

Oxygen sensor alarm

The oxygen sensor alarm will be triggered if the percentage of oxygen in the room is too low or too high. In both cases, serious injuries or death can occur.

<fc #ff0000>If the alarm goes off</fc>

LN2 leak

<fc #ff0000>If the dewar is leaking and there is liquid nitrogen on the floor,</fc>

Otherwise

Personal Protective Equipment

While manipulating the dewar and testing using liquid nitrogen,

*Always wear:

* Do not wear metal jewelry or watches

* Place the oxygen sensor close to you or on you.

Before thermomechanical testing

The dewar: The Airgas dewars generally have metal plaques on each valve indicating which is the vent and which is the liquid.